Fluid Structure Interaction (FSI) of endovascular repair (EVAR) of AAA
The
local dilation of the infrarenal aorta, termed an abdominal aortic
aneurysm (AAA), affects 2-4% of the adult population in the United
States and may eventually lead to rupture, an event currently ranked
the 15th leading cause of death. For those
patients whose AAAs grow to a critical diameter, it becomes necessary
to treat the aneurysm either by open surgery and prosthetic replacement
or by endovascular repair (EVAR). EVAR was introduced in the early
1990’s and represents a significant improvement over open
surgery. This procedure involves inserting an endovascular graft into
the aneurysm to prevent pressurization of the AAA cavity. As compared
to open surgery, EVAR results in lower blood loss and a shorter
hospital stay for the patient. Despite the promising use of EVAR in
AAA, this procedure is prone to several failure mechanisms including
but not limited to graft migration and endoleak.
Endoleak is defined as the flow of
blood into the AAA sac after EVAR. There are several types of endoleak,
each defined by the mechanism through which blood enters the AAA sac.
Type I endoleak occurs when there is an ineffective seal between the
endovascular graft and the AAA neck. While type I endoleak is a rare
complication after EVAR, it is associated with a high number of
reinterventions and potentially serious consequences.
The purpose of this study is to evaluate the influences of several clinically relevant variables on the occurrence of failure (e.g., type I endoleak). In addition, the geometric and material properties of the graft will be modified in order to minimize the likelihood of leakage and migration. In order to accomplish this, a finite element model of a grafted AAA will be analyzed using LS-DYNA, a transient dynamic finite element code. Direct coupling between blood flow and the grafted artery is simulated in LS-DYNA with a multi-material formulation. This coupling will allow for accurate predictions of how blood flow will affect the performance of endografts in a variety of AAA configurations.